TUR, TAR and guard-band calculator

SCQUALITYGUILD.COM/TOOLS/TUR-TAR-GUARD-BAND-PFA.HTML

Enter the tolerance, the calibration uncertainty and how often instruments like this one arrive in tolerance. Get the test uncertainty ratio, the acceptance limits for several guard-band methods, and the probability of a false accept and a false reject for each.

ON THE EXAM · CCT IV.A, IV.C, CQE IV.E · FREE · NOTHING YOU TYPE LEAVES YOUR BROWSER

Q-BotClick for an explanation of this tool.

For educational use. This page is a study aid. The method behind it has far more depth and detail than one page and one tool can cover, and what you see here is not everything the method offers or everything this site offers. For full understanding and application, use the governing standards and texts, and seek further education, training or qualified advice. Results depend on the data you enter, so check them before relying on them.

TUR and TAR

The test uncertainty ratio compares the tolerance being checked with the uncertainty of the check. ANSI/NCSL Z540.3 defines it as the tolerance span divided by twice the expanded uncertainty of the calibration process at about 95% (k = 2): TUR = (upper limit − lower limit) / 2U. For a symmetric tolerance ±T that is T/U.

The older test accuracy ratio compares the tolerance with the accuracy specification of the reference standard alone. TAR is easier to work out but leaves out resolution, repeatability, the environment and the method, so it nearly always looks better than the TUR. The common 4:1 rule of thumb is a TUR requirement.

False accept and false reject

Every reading is off by some amount, so near a limit a calibration can get the answer wrong in two ways.

Both depend on the uncertainty and on how the instruments are spread: a population that is nearly always well inside tolerance gives few false accepts whatever the TUR. Here the true values are taken as normal, with the spread set by the in-tolerance probability (end-of-period reliability, EOPR) from your calibration history. The measurement errors are normal with standard deviation U/k. PFA and PFR are the double integrals over the regions where truth and decision disagree, worked out numerically. Z540.3 requires PFA of 2% or less, or, where PFA is not calculated, a TUR of at least 4:1.

Guard bands

A guard band moves the acceptance limit A inside the tolerance limit T, so a reading must be further from the limit to pass. It lowers PFA and raises PFR. The methods compared here:

The trade-off chart shows the cost: every step inward buys a lower PFA with a higher PFR.

Where U comes from

U is the expanded uncertainty of the calibration at this test point, not of the instrument being calibrated. Build it from its sources with the measurement uncertainty budget.

On the exam

CCT IV.A lists guard-banding, PFR (Type I error), PFA (Type II error), TUR, TAR and percent of tolerance among the uncertainty terms to apply. IV.C covers combined and expanded uncertainty, coverage factors and the statement of conformity with its decision rule. CQE IV.E covers metrology, including measurement error and its sources. Expect to compute a TUR or TAR, apply a guard band, and say which way a guard band moves each risk.

Other tools and templates